M2000 Series Data Recorders

Specification

Mainframe  
Number of Channels: 8 or 16, can be increased by mix of analog and l6ch digital boards.
Recording Technique: Thermal array, direct writing, high resolution black on white permanent record.
Paper Loading: Drop-in roll, no threading.
Paper Size:
(Roll)
M2000 Width up to 216mm (8.5) Length 30m (100’) or 60m (200’)
M2200 Width up to 155mm (6”) Length 40m (130’)
Record Length: Selectable in seven steps from 10cm to continuous.
Chart Speeds: 24 speeds from 0.5mm/hr to l0m/sec in real time and/or storage mode.
Channel Configuration: All channels with full paper width overlap or switch selectable to 4 or 8 parallel tracks.
Bandwidth: DC to 10kHz (Up to 80kHz sampling/channel).
Event Marker: Underscores text for duration of event. Operated from front panel or remote contact closure.
Grid Lines: Selectable from front panel. Off, 1cm grid, 1 cm grid plus 2mm fine grid.
Memory: 128 Kbytes per channel, (12 bit ADC), standard. Extendable to 512 Kbytes.
Trace Viewing : Integral, high resolution LED array provides visual presentation of all channels.
Test Annotation: Time, date, run number etc., automatically printed along paper edge. (Test details can be added via the system computer).
Channel Identification: Channel number printed at regular intervals alongside data.
Memory Contents: Front panel bargraph status indicator.
Dimensions (M2000): Height 177mm (7”). Width 400mm
(15.75”). Depth 460mm (18”). Excluding feet and connectors.
Weight 15kg.
Dimensions (M2200): Height 176mm (7”). Width 280mm
(11”). Depth 350mm (14”). Excluding feet and connectors.
Weight 12kg.
Power Supply: 115/230 volts, 45-440Hz (by voltage selector), plus optional integral inverter for 1 2/24V d.c.
Consumption: 8OVA (M2200-8). 100VA (M2000-16).
Signal Conditioning  
Signal Inputs: Any type of transducer plus analog or digital signals from microvolts through to 500V rms.
Digital Monitor: 31/2 digit dual range auto-polarity display for reading channel number, voltage output, calibration input and trace start point.
Transducer Supply: Adjustable in lv steps between 3 and 12V D.C. Max current: 1A.
Calibrator: Continuously variable calibration from ±1.000mV to ±10.OOV by front panel control and digital indicator.
Voltage Outputs: All amplifiers/signal conditioners can have optional voltage outputs via rear mounted connectors to drive tape recorders, oscilloscopes and data acquisition systems.
Transient Capture  
Trigger Polarity: Positive, negative or both (Switch selectable).
Trigger Allocation: Single or multi-channel in any mix with visual indication of trigger state on all channels.
Pre-Event: Trigger point can be set at five separate stations by selector switch. (Start, 25%, 50%, 75% or End of record).
Trigger Mode: Manual or alternative automatic monitoring where unattended operation is desirable, (i.e. “baby-sit” mode).
Timebase Expansion: Data can be replayed as captured, or expanded for detailed analysis by a factor of up to 50:1.
OPTIONAL EXTRAS  
 

Ruggedised Transit Case 19” Rackmount Adaptor Hand Held Remote Controller 12/24v D.C. Operation Microlog 2000 Data Acquisition Software